bowie & Dick tests
With reference to the European Standard EN 285, the Bowie & Dick Test is a mandatory test for steam sterilisers...
With Famos we are able to provide Bowie & Dick testing equipment to complement porous and hollow loads
as well as testing for Class B bench top sterilisers.
This functionality test for steam sterilisers needs to be done at the beginning of the working day
and following a successful pass of the Bowie & Dick Test the steam steriliser can be released for use.
The PCD units of the the helix tests represents a greater test challenge for steam sterilisers and the sterilisation process as the indicator cards are sited within a retaining device attached to a narrow lumen tube.
The Bowie & Dick Test Pack and Bowie & Dick Test Set comply to EN ISO 11140-1 (type 2) as well as EN ISO 11140-4
(Click on the images below to follow the links)
With Famos we are able to provide Bowie & Dick testing equipment to complement porous and hollow loads
as well as testing for Class B bench top sterilisers.
This functionality test for steam sterilisers needs to be done at the beginning of the working day
and following a successful pass of the Bowie & Dick Test the steam steriliser can be released for use.
The PCD units of the the helix tests represents a greater test challenge for steam sterilisers and the sterilisation process as the indicator cards are sited within a retaining device attached to a narrow lumen tube.
The Bowie & Dick Test Pack and Bowie & Dick Test Set comply to EN ISO 11140-1 (type 2) as well as EN ISO 11140-4
(Click on the images below to follow the links)